Tier 3 only, Tier 1, 2 & 3 – Activeforever RemRest 901 CPAP Machine User Manual
Page 10

VIEW OR SET THE DYNAMIC CALIBRATION:
This mode allows you to set a resistance compensation factor
determined by the type of interface device used. When using a
nasal mask or a full face mask, press the up or down arrows
to select “r nr”. If using a nasal device such as the Nasal-Aire I or
II or the Spiritus, press the up or down arrows to select
“r r”, the blower function will compensate for device restriction.
Once the desired setting is reached, press the mode button to save
the setting.
For Tier 2, the display is now back to prescription pressure and
you may review all mode settings or turn the power off to exit
technician mode.
For Tier 3, the display will show “LOAd” only. This is for
factory use only. You may push mode to get back to prescription
pressure or turn the power off to exit technician mode.
WARNING:
The Tier 2 and 3 units are shipped with a dust cover on the DB9
connector. This cap should be left on except when the connector is
in use.
Tier 3 Only:
AUTO MASK CHANGE REMINDER INDICATOR & RESET:
The “TIME” Icon on the LCD flashes after 90 days total time.
To reset, press and hold the mode and down arrow button for
3 seconds. The “Time” Icon goes off and the 90 day time interval
re-starts. This feature can only be disabled with an external
command via the external software and DB9 connector.
Tier 1, 2 & 3:
To exit out of the technician mode on all Tier models, press the on/off
button at any time. The light will turn off.
10 RemRest