NOTE: Tier 1 model does not incorporate automatic altitude

compensation. In order to insure that this device provides the

expected pressures, the following adjustments to the pressure

setting must be made to the device:

For an increase in altitude, please multiply the prescription

pressure setting by the correction factor found in the table.

For a decrease in altitude, please divide the prescription pressure

setting by the correction factor found in the table.

RemRest 5

Altitude Change (+/- Meter)

457

914

1372

1829

2286

2438

Altitude Change (+/- Feet)

1500 3000 4500

6000

7500

8000

Correction Factor

1.05

1.11

1.16

1 21

1 27

1 32