Activeforever RemRest 901 CPAP Machine User Manual
Page 4

TO SET THE PRESCRIPTION PRESSURE
(BLINK ONCE/OFF 3 SECONDS):
The default setting is 10cmH2O. Using a manometer and pushing
the up and down arrows, adjust the pressure to the prescribed
setting.
TO SET THE RAMP STARTING PRESSURE
(BLINK TWICE/OFF 3 SECONDS):
The default setting for the starting ramp pressure is 3.5 cmH2O.
To enter the starting ramp set mode, press the white dot once. The
white dot is located between the on/off button and ramp button.
Using a manometer and pushing the up and down arrows, adjust
the pressure to the prescribed setting.
TO SET THE RAMP TIME
(LIGHT BLINKS 3 TIMES/OFF 3 SECONDS):
To enter the set ramp time mode, press the white dot a second
time. Press the up and down arrow to set the ramp time. The ramp
light will blink based on the ramp delay time. Push the up arrow
again to return to the prescription pressure.
Before exiting the technician mode, remember to always press the
white dot button after any change to lock the new settings into
memory. To exit the Technician Mode at any point, press the on/off
button. Both the on/off and ramp lights will go out.