Accu-Chek Advantage User Manual
Page 56
109
108
IF...
the following screen is displayed.
REASON
The meter has failed its internal system check.
ACTION
Remove the Chek
™
Strip. Turn the meter off and
repeat the glucose control check. See
Glucose
Control Testing
(page 27).
If this error display still
appears, testing cannot take place. Call Accu-
Chek
SM
Customer Care at 1-800-858-8072.
IF...
the following screen is displayed.
REASON
The test strip may be damaged or the test was not
performed correctly.
ACTION
If this display appears
before blood is placed on the
strip, remove the test strip and reinsert. If this
display appears
after blood is applied to the strip,
there is an incorrect amount of blood on the strip.
Repeat the test with a new test strip.
Troubleshooting Guide