Accu-Chek Advantage User Manual
Page 54
105
104
IF...
the following screen is displayed.
REASON
The temperature is outside the operating range of
the test strips (57°-104°F or 14°-40°C).
ACTION
The result you have obtained may not be
accurate. A temperature symbol will appear below
your result. Move to an area between 57° and
104°F, wait 5 minutes and repeat the test. Do not
artificially heat or cool the meter.
▲ Example result
Blue circle placed
over screen captures
prints PMS 287.
IF...
the following screen is displayed.
REASON
The meter is turned on and the memory is full with
100 values. MEM flashes.
ACTION
If a new result is added, the oldest result is deleted.
▲ Example code number
Troubleshooting Guide
362-13701 AdvNonSAMa 9/13/02 8:20 AM Page 104