Leak test temperature, Co dilution, A. test a - one minute let-by test – RIDGID FG-100 User Manual
Page 8

7
Change page on display pressing
6. View the gas analysis values from the 6 different pages
7. Print the report
Report includes all analysis data, the draft and
smoke index values if performed.
All measured and calculated values can be printed.
Press:
Zoom in and out by pressing.
Leak Test
Temperature
Tflow/Treturn can be measured using the standard thermocouple input.
Connect the Tc K to the instrument.
Press
to exit to the previous menu.
F2
Select Flow/Return parameter for changing, pressing
Press
“Copy” to copy the actual temperature to the Flow/Return parameter.
F1
Press:
Select and confirm the “
” option.
TEMPERATURE
Execute the steps above for other parameter in the same way.
Press
to print the values.
ATTENTION
THE USAGE OF A PRESSURE HIGHER THEN ±300 HPA WILL DAMAGE
THE THE PRESSURE SENSOR OF THE INSTRUMENT PERMANENTLY.
Two pressure network leak test can be performed using a pressure decay
program.
A. Test A - one minute Let-by test
Press:
Select and confirm the “
” option.
TIGHTNESS
Select and confirm the “
” option.
Test type
Change and confirm test type to “Test A” for one minute let-by test.
Press the “ZERO” function (F1) to zeroing the pressure sensor.
Connect the pressure input
to the gas leak test kit (cat. 26028).
Connect the kit to the gas network. If required, use the rubber blanking
plugs for connecting.
Open the valve on "T" connector.
Pressurize the network at the testing pressure.
Close the valve on "T" connector.
Select the “START” option pressing
P1
–
TEMPERATURE
Temper.:
40.0 °C
COPY
ESCAPE
Flow...:
40.0 °C
Return.:
--.- °C
Differ.:
--.- °C
–
SELECT TEST
0.3 hPa
Test Type : A
START TEST
ZERO
ESCAPE
CO Dilution
The CO dilution system can be enabled or disabled from the Configuration
page.
Press the
key to open the configuration menu.
Press "
"
Select and confirm the "Dilution" option
Select the desired option:
(the dilution automatically start when the CO
concentration is higher than 2000 ppm)
(the dilution is always enabled)
(the dilution is always disabled)
NEXT
AUTO
ON
OFF
–
CONFIGURATION P1
--- FUEL ---
> LPG
CO SENS.
Includ.
FUEL PARAMETERS
NEXT
ESCAPE
–
CONFIGURATION P2
ENGINEER UNITS
ALARMS
O2 REFERENCES
Dilution AUTO
NEXT
ESCAPE
NOTE
The supplied syringe with the optional leak test kit is intended for
applications in Italy and Germany only.