Leak test temperature, Co dilution, A. test a - one minute let-by test – RIDGID FG-100 User Manual

Page 8

background image

7

Change page on display pressing

6. View the gas analysis values from the 6 different pages

7. Print the report

Report includes all analysis data, the draft and
smoke index values if performed.

All measured and calculated values can be printed.
Press:

Zoom in and out by pressing.

Leak Test

Temperature

Tflow/Treturn can be measured using the standard thermocouple input.

Connect the Tc K to the instrument.

Press

to exit to the previous menu.

F2

Select Flow/Return parameter for changing, pressing

Press

“Copy” to copy the actual temperature to the Flow/Return parameter.

F1

Press:

Select and confirm the “

” option.

TEMPERATURE

Execute the steps above for other parameter in the same way.

Press

to print the values.

ATTENTION

THE USAGE OF A PRESSURE HIGHER THEN ±300 HPA WILL DAMAGE

THE THE PRESSURE SENSOR OF THE INSTRUMENT PERMANENTLY.

Two pressure network leak test can be performed using a pressure decay
program.

A. Test A - one minute Let-by test

Press:

Select and confirm the “

” option.

TIGHTNESS

Select and confirm the “

” option.

Test type

Change and confirm test type to “Test A” for one minute let-by test.

Press the “ZERO” function (F1) to zeroing the pressure sensor.

Connect the pressure input

to the gas leak test kit (cat. 26028).

Connect the kit to the gas network. If required, use the rubber blanking
plugs for connecting.

Open the valve on "T" connector.

Pressurize the network at the testing pressure.

Close the valve on "T" connector.

Select the “START” option pressing

P1

TEMPERATURE

Temper.:

40.0 °C

COPY

ESCAPE

Flow...:

40.0 °C

Return.:

--.- °C

Differ.:

--.- °C

SELECT TEST

0.3 hPa

Test Type : A

START TEST

ZERO

ESCAPE

CO Dilution

The CO dilution system can be enabled or disabled from the Configuration
page.

Press the

key to open the configuration menu.

Press "

"

Select and confirm the "Dilution" option

Select the desired option:

(the dilution automatically start when the CO

concentration is higher than 2000 ppm)

(the dilution is always enabled)

(the dilution is always disabled)

NEXT

AUTO

ON

OFF

CONFIGURATION P1

--- FUEL ---

> LPG

CO SENS.

Includ.

FUEL PARAMETERS

NEXT

ESCAPE

CONFIGURATION P2

ENGINEER UNITS

ALARMS

O2 REFERENCES

Dilution AUTO

NEXT

ESCAPE

NOTE

The supplied syringe with the optional leak test kit is intended for

applications in Italy and Germany only.