beautypg.com

Reprocessing the mask between patients, Technical specifications – ResMed Mirage Liberty User Manual

Page 24

background image

24

Reprocessing the mask between patients

The mask should be reprocessed when used between patients. Cleaning,
disinfection and sterilisation instructions are available from the ResMed website,

www.resmed.com/masks/sterilization. If you do not have internet access,
please contact your ResMed representative.

Technical specifications

Pressure-flow curve The mask contains passive venting to protect against

rebreathing. As a result of manufacturing variations, the vent flow rate may vary.

Vent flo

w rat

e (L/min)

0

20

40

60

80

2

4

6

8

10

12

14

16

18

20

22

24

26

28

30

32

34

36

38

40

Quattro Air/ Quattro Air for Her

Quattro

Quattro FX/ Quattro FX for Her

Liberty

Mask pressure (cm H

2

O)

Quattro Air/

Quattro Air

for Her

Quattro FX/

Quattro FX

for Her

Quattro

Liberty

Pressure (cm H

2

O) Flow (L/Min)

3

19

4

22

22

22

22

8

32

32

32

32

Quattro Air/

Quattro Air

for Her

Quattro FX/

Quattro FX

for Her

Quattro

Liberty

Pressure (cm H

2

O) Flow (L/Min)

10

37

37

37

37

12

41

41

41

41

16

48

48

48

48

20

54

54

54

54

24

60

60

60

28

66

66

66

30

69

69

69

32

72

72

36

77

77

40

82

82

Deadspace: Empty volume of mask to swivel. The deadspace of the mask varies
according to cushion size.

222 mL
(Large)

243 mL
(Large)

240 mL
(Large)

151 mL
(Large)*

* Empty volume of mask to elbow.
Therapy pressure (cm H

2

O)

3 - 40

4 - 30

4 - 40

4 - 20

Resistance: Drop in pressure measured (nominal) (cm H

2

O)

at 50 L/min

0.2

0.1

0.1

0.1

at 100 L/min

0.9

0.5

0.4

0.5

Inspiratory and Expiratory Resistance (cm H

2

O)

Inspiration

0.6

0.7

0.8

0.7

Expiration

1.0

0.7

0.8

0.8