8. Reference to standards

Device standard:

Device corresponds to the requirements of the

EN1060-1 /-2; ANSI / AAMI SP09

This device complies with the requirements of the Medical Device Directive 93/42/EEC.

9. www.microlife.com

Detailed user information about our products as well as services can be found at www.microlife.com

10. Technical specifications

Weight:

360 g

Size:

175 x 70 x 103 mm

Storage temperature:

­20 °C to +70 °C; 85 % relative humidity maximum

Operation temperature: 0 °C to 46 °C

Measuring range:

0 to 300 mmHg

Measuring resolution: 2 mmHg

Accuracy:

within ± 3 mmHg in 18 °C to 33 °C;

within ± 6 mmHg in 34 °C to 46 °C

Inflation source:

a volume of at least 200cc to a pressure of 300 mmHg in 4 to 10 sec

Air leakage:

<± 4 mmHg/min

Hysteresis error:

within 0 mmHg to 4 mmHg

Accessories:

1. cuff (with arm circumference of 25 - 40 cm) with inlaid bladder

2. bulb and valve

3. soft bag

Technical alterations reserved.