Microlife AG1-10 User Manual
Page 12
9
☞
Further information
Measurements should not occur soon after each other, since otherwise, the results will be
falsified. Wait therefore for several minutes in a relaxed position, sitting or lying, before you
repeat a measurement.
5. Other possible malfunctions and their elimination
If problems occur when using the device, the following points should be checked and if
necessary, the corresponding measures are to be taken:
Malfunction
The sound transmission is
poor, distorted or there is
extraneous noise.
The pressure does not rise
although the bulb is pumping.
The deflation rate can not
be set to 2 - 3 mmHg/sec. by
adjusting the air release valve.
Pointer is not at 0 +/- 3 mmHg
at rest.
☞
FURTHER INFORMATION
The level of blood pressure is subject to fluctuations even with healthy people. Important thereby
is, that
comparable measurements always require the same conditions (rest condition)!
If, in spite of observing all these factors, the fluctuations are more than 15 mmHg, and/or you
Remedy
1. Check the earpieces if they are plugged or cracked. If not,
make sure they do not fit poorly as worn.
2. Check the tube if it is broken or twisted.
3. Check the bell and diaphragm of chestpiece if there is any crack.
4. Make sure the chestpiece is in proper contact with skin and
over brachial artery during measuring. Clean or replace any
defective parts if found to avoid inaccurate reading.
1. Make sure that the valve is closed.
2. Make sure the cuff is properly connected to bulb and manometer.
3. Check if the cuff, tube and bulb is leaky. Replace the defective
parts if any.
• Disassemble the valve from bulb to check if there is any blockage
in the airway of the valve. Clean the blockage and try again. If
it still does not work, replace it to avoid inaccurate reading.
1. Make sure that the valve is open for zero check.
2. If still more than 3 mmHg deviation, contact your dealer to
recalibrate the manometer.