beautypg.com

Appendix, A-1 appendix appendix appendix appendix appendix, A.1 specifications – Teledyne 3160 User Manual

Page 59

background image

A-1

Appendix

Appendix

Appendix

Appendix

Appendix

TELEDYNE ANALYTICAL INSTRUMENTS

Appendix

A.1

Specifications

(Subject to change without notice.)

Ranges:

(Six ranges with AutoRanging on all ranges)

1:

0–1 part per million (ppm) O

2

2:

0–10 ppm O

2

3:

0–100 ppm O

2

4:

0–1000 ppm O

2

5:

0–1 % O

2

6:

0–25 % O

2

Accuracy:

±2% of full scale when calibrated and operated at the same constant
temperature and pressure.*
±5% of full scale over the operating temperature range once tempera-
ture equilibrium is reached.*

*

Additional 0.04 ppm on 0-1 ppm range, generated by internal scrub-
ber (which is necessary to maintain these specifications).

Response Time: 90 % of final reading in less than 61 seconds at 25 °C

Operating Temperature:0-50°C (32

-122 °F)

Analog Signal Outputs:

For percent of full scale indication:

one 0–1 V dc non-isolated signal
one 4–20 mA dc isolated signal

For range identification:

one 0–1 V dc non-isolated signal
one 4–20 mA dc isolated signal

Digital Data Lines:

One RS-232C serial interface for 2-way communications
with separate host computer—for remote monitoring and
control of all functions.

Alarms:

Five user-programmable absolute-reading alarm setpoints,
with Form C, SPDT relays (3A resistive).

This manual is related to the following products: