TELEDYNE ANALYTICAL INSTRUMENTS
1. Enter the Calibration mode from the Analyze mode by
pressing SPAN.
2. Press ACK if zeroing has already been done. Make sure
to zero before calibrating to ensure correct O
2
readings.
3. If the span gas tank is not connected to the span port,
connect it now and press ACK.
4. Enter the span gas concentration using the UP and
DOWN keys, and press ACCPT.
5. Set the flow rate to 2 SCFH and press ACK.
6. Select manual or automatic sensor settling detection by
pressing MAN or AUTO. Automatic sensor settling de-
tection lets the system determine when the sensor has
stabilized.