TELEDYNE ANALYTICAL INSTRUMENTS

1. Enter set-up from the Analyze mode by pressing

SETUP.
If only one cell is installed, the TEST and INSTL op-
tions on the next screen are omitted.

2. Press TEST to test the alternate sensor or INSTL to in-

stall a new sensor.

3. If TEST was chosen in step 2, the active sensor is dis-

played and you are directed to press ACK to test the al-
ternate sensor.

If INSTL was chosen in step 2, go to step 5.

4. After pressing ACK in step 3, the alternate sensor is

flushed for 10 minutes while the measured O

2

level is

displayed. The 10 minute flush may be restarted by
pressing RESET, or aborted by pressing SKIP. Go to
step 8.

5. If INSTL was chosen in step 3, the screen will display

the active sensor. Press A or B for the sensor you are
installing.

6. Install the sensor (either A or B from step 5) into the

sensor block within two minutes, and press ACK.