Installation 3

Installation 3

Installation 3

Installation 3

Installation 3

3-9

TELEDYNE ANALYTICAL INSTRUMENTS

WARNING:

Power cord must be disconnected before performing any
voltage selection!

To change from North American to European fuses, perform the following:

1. Use a small blade screwdriver to open the cover.

2. Loosen the Phillips screw two turns and remove the fuse block by sliding it

away from the screw and up from the pedestal.

3. Remove the North American fuse and turn the fuse block over.

4. Replace the fuse with European fuses. Two European fuses are required,

but the lower one may be replaced with a dummy fuse, or jumper bar (see
Figure 3–6).

Figure 3–6: Fuse Replacement

5. Slide the fuse block back on the Phillips screw and pedestal.

6. Tighten the Phillips screw and replace the cover.

NOTE: The fuses that go into the housing first are the active set.

European 1.6 Amp T Fuses

5 x 20 mm

North American 3 A T Fuses