Table 4-1: Front Panel Controls and Indicators
Number
Control/Indicator
Description
Section
4
CURRENT display
4 digit, 7 segment LED display. Normally displays the Out-
put Current. When the PREV button is pressed, the display
indicates the programmed setting of Output Current.
5
CURRENT indicator
Green LED, lights for Constant-Current mode operation
6
CURRENT control
High resolution rotary encoder for adjusting the Output Cur-
rent.
Auxiliary Function:
1. Selects the Baud-Rate of communication port.
2. Set desired mode of the Master Power Supply in Ad-
vanced parallel operation.
5.2.2
7.2.4
5.15.2
7
OUT button
Main function: Output ON/OFF control. Press OUT to set
the output On or Off. Press to reset and turn On the output
after OVP or FOLD alarm events have occurred. Auxiliary
function: Selects between “Safe-Start” and “Auto-Restart”
modes. Press and hold OUT button to toggle between
“Safe-Start” and “Auto-Restart”. The VOLT display will cycle
between “SAF” and “AUT”. Releasing the OUT button while
one of the modes is displayed, selects that mode.
5.6
5.11
8
OUT indicator
Green LED, lights when the DC output is enabled.
9
REM/LOC button
Main function: Go to local. Press REM/LOC to put the unit
into Local mode (REM/LOC button is disabled at Local
Lockout mode).
Auxiliary function: Address and Baud Rate setting. Press
and hold REM/LOC for 3 sec. to set the Address with the
VOLTAGE encoder and the Baud Rate with the CURRENT
encoder.
7.2.5
7.2.2
7.2.4
10
REM/LOC indicator
Green LED, lights when the unit is in Remote mode.
11
FOLD button
Foldback protection control.
-Press FOLD to set Foldback protection to On.
-To release Foldback alarm even, press OUT to enable the
output and re-arm the protection.
-Press FOLD again to cancel the Foldback protection.
5.5
12
FOLD indicator
Green LED, lights when Foldback protection is On.
13
OVP/UVL button
Over Voltage Protection and Under Voltage limit setting.
-Press once to set OVP using VOLTAGE encoder (the cur-
rent display shows “OUP”)
-Press again to set the UVL using VOLTAGE encoder (the
current display shows “UUL”).
5.3
5.4