beautypg.com

Adsorber replacement, Figure 5-1: disconnecting self sealing couplings, Install the compressor rear panel – Olicom 9600 User Manual

Page 45: Adsorber replacement -2, Ti cryogenics

background image

Maintenance

5-2

P/N 8040540

C

HELIX TECHNOLOGY CORPORATION

-

TI

CRYOGENICS

Adsorber Replacement

Use the following procedure to change the adsorber every three years.

1. Set the System Circuit Breaker, on the rear of the 9600

Compressor, to the OFF position.

2. Remove the 4 screws which secure the rear panel to the Compres-

sor and remove the rear panel.

NOTE: Use two wrenches in Step 3 to prevent loosening the body of the
coupling.

3. Using a 1-3/16 in. wrench, and a 1-1/8 in. wrench, as shown in Fig-

ure 5-1, disconnect the two self sealing coupling connectors quickly
to minimize helium leakage.

Figure 5-1: Disconnecting Self Sealing Couplings

4. Using a 7/16 in. (11mm) wrench, remove the adsorber mounting

bolt as shown in Figure 5-2.

5. Move the adsorber from under the mounting tabs in the base as

shown in Figure 5-2 and remove the adsorber from the Compressor.

6. Install the replacement adsorber under the mounting tabs and

secure it into place with the bolt removed during Step 4.

7. Using two wrenches as shown in Figure 5-1, connect the two self

sealing couplings quickly to minimize helium leakage.

8. Install the Compressor rear panel.

This 1 3/16 in. wrench
is used to turn the self
sealing coupling
connector.

This 1 1/8 in. wrench
holds the coupling in a
stationary position.

Side View

To Tighten Coupling Halves

Hold Here

Turn Here

Hold Here

Turn Here

To Loosen Coupling Halves