Omega 24 User Manual
Page 7
BATTERY LIFE, CONTINUOUS: 10 hrs. typical, alkaline; 3 hrs typical, Ni-Cd (rechargeable).
BATTERY INDICATOR: Display indicates BAT when less than 10% of life remains.
DIMENSIONS, WEIGHT: 7.0" x 2.9 x 1.1". Net weight 12 oz.
UNIQUE SPECIFICATIONS
NOTE: The specifications below are specific to the model(s) as identified by model number. The
difference specifications are in addition to the OVERALL SERIES SPECIFICATIONS found on
pages 2-5.
• MODEL CL23A/CL23EC:
SENSOR SELECTION: K, J, T Thermocouples
ACCURACY: (18°C to 28°C ambient, 2 years): ±0.5°F (rdg > -50°F).
±0.04% rdg (rdg > 1250°F)
±1.0°F (rdg < -50°F)
TEMPERATURE COEFFICIENT: 18°C to 28°C; included in accuracy specifications.
From 0°C to 18°C, and 28°C to 50°C; less than ±(0.006% rdg +0.03°C)/°C.
RAMP FUNCTION: Not Included.
BATTERY LIFE, CONTINUOUS: 16 hrs typical, alkaline; 4 hrs typical, Ni-Cd (rechargeable).
• MODEL CL24:
SENSOR SELECTION: K, J, T, E Thermocouples
ACCURACY: (18°C to 28°C ambient, 2 years): ±0.5°F (rdg > -50°F).
±0.04% rdg (rdg > 1250°F)
±1.0°F (rdg <-50°F)
TEMPERATURE COEFFICIENT: 18°C to 28°C; included in accuracy specifications.
From 0°C to 18°C, and 28°C to 50°C; less than ±(0.006% rdg +0.03°C)/°C.
RAMP FUNCTION: Not Included.
BATTERY LIFE, CONTINUOUS: 16 hrs typical, alkaline; 4 hrs typical, Ni-Cd (rechargeable).
• MODEL CL25/CL25EC:
SENSOR SELECTION: K, J, T, E, N, B, R, S, G, C, D Thermocouples
RAMP FUNCTION: Included.
BATTERY LIFE, CONTINUOUS: 16 hrs typical, alkaline; 4 hrs typical, Ni-Cd (rechargeable)
• MODEL CL26:
SENSOR SELECTION: K, J, T, E Thermocouples
100Ω Platinum RTD
2252Ω Thermistor
1,000Ω/100,000Ω Resistance
5