SERVICE INFORMATION

WARNING

All service information is intended for qualified electronic mainte-

nance

personnel

only.


1. CALIBRATION PROCEDURES

This is a two-part procedure. Products with thermocouple functions only, follow the
Thermocouple Calibration procedure below. Products with resistance functions (RTD,
thermistor, ohms) follow first the Thermocouple Calibration procedure, then the Resistance
Calibration procedure.

A. THERMOCOUPLE CALIBRATION

Test Equipment Required:

1. Thermocouple simulator (OMEGA CL521, or equivalent) calibrated to ITS-90.
2. Microvolt DMM (OMEGA OM7551, or equivalent)
3. Calibration cables, per Figures 11, 12, 13.


NOTE: The copper constantan cables used in Figures 12 and 13

need to be calibrated. Voltage errors in these cables should be com-

pensated

at

the

simulator.

4. Calibration cover (OMEGA 820-307-4).

Ambient Conditions:


Units should be calibrated at an ambient temperature of 23°C ±1°C, with relative humidity less

than 80%.

Preparation for Calibration:

1. Remove battery from unit under test (U.U.T.). Refer to Battery

Installation/Replacement section in the manual.

2. Remove bottom-cover from U.U.T. Refer to Disassembly Instructions.

3. Remove calibration-jumpers J1 and J2 from printed-circuit board.


NOTE: Removal of J1, J2 causes partial loss of previously stored

calibration data. Do not remove these jumpers unless recalibration

is intended.