23. Select Type-E thermocouple.

24. Turn off U.U.T.


25. Replace T calibration-cable with E calibration-cable.

26. Turn on U.U.T. Allow 2-3 minutes for thermal stabilization.

27. Apply 32°F (Type E). Adjust P1 for a reading of 32.0°F (exclusive of
noise).

28. Hit OPR key to store E-meter zero-adjustment.

29. Select Type-K thermocouple.


30. Select CALIB mode.

31. Turn off U.U.T.

32. Connect U.U.T. to test-equipment per Figure 3 (page 43) using the K calibration-
cable.

33. Turn-on U.U.T. and allow 2-3 minutes, for thermal stabilization

34. Set both the U.U.T. and TC-simulator to output 32.0°F (Type K). Adjust
P2 for a null-meter reading of 0±3mv.

35. Set both the U.U.T. and T-simulator to output 2400°F (Type K). Adjust
P41 for a null-meter reading of 0±3mV.

NOTE: Steps 34 and 35 are interactive. Repeat each as neces-

sary to get the desired null readings.


36. Hit OPR key to store K-calib zero adjustment.

37. Select J thermocouple.

38. Turn off U.U.T.

39. Replace K calibration-cable with J calibration-cable.


40. Turn on U.U.T. Allow 2-3 minutes for thermal stabilization.

41. Set both the U.U.T. and TC-simulator to output 32.0°F (Type J). Adjust
P1 for a null-meter reading of 0±3mV.


42. Hit OPR Key to store J-calib zero adjustment.

43. Select Type-T thermocouple.

44. Turn off U.U.T.

41