3. Select the calibrator mode ("CALIB" on display) by pressing the
CALIB/METER function button.
4. Select type K thermocouple ("K" on display) by pressing the SENSOR
SELECT function button until type K is reached.
5. Select the tenth degree display with a digit appearing after the decimal
point on the display) by pressing the 0.0.1 function button.
6. Select °F ("°F" on display) by pressing the °F °C function button.
7. Attach the Alloy Mode Calibrator Verification Cables per Fig. 13.
8. Press the CHANGE function button and the numerical display will blink
and "NUM" will appear on the display.
9. Press 3 2 0 buttons and then press the ENTER function button. "NUM"
will disappear and "32.0 °F" will appear, not blinking.
10. Adjust the Thermocouple Temperature Simulator to 32.0 °F.
Note: The reading on the voltmeter should read null. 0.000 mV. Adjusting
the U.U.T. to any temperature and adjusting the Temperature Simulator to
the same temperature should result in a null, or 0.000 mV on the volt-
meter.
11. Disconnect the cables from the equipment and UUT and reconfigure the set-up
per figure 12.
12. Select the meter mode by pressing the CALIB/METER function button on the
U.U.T. and ensure the U.U.T. is still set for °F and type “K” thermocouple.
13. Set the thermocouple simulator to type “K” and an output of 32.0°F. The
U.U.T. should read 32.0°F.
14. Set the thermocouple simulator to 2500°F. The U.U.T. should read 2500°F.
15. Set the thermocouple simulator to - 320°F. The U.U.T. should read -320°F.
16. Remove the Alloy Mode Calibrator Verification Cable from all equipment.
17. The Alloy Mode verification is now complete. Press the OFF function but-
ton to turn the unit off.
37