• With J1 removed, the digital-filters in the U.U.T. are dis-

abled. This is to speed up U.U.T. response to calibra-

tion adjustments. In this mode ±0.1 °F of reading noise

is normal.


• Do not deviate from the calibration adjustment sequence

that follows. This will ensure that zero adjustments for each

TC type will be stored in the appropriate memory locations.


8. Apply 32°F (Type K) to U.U.T. Adjust potentiometer P1 for a reading of
32.0°F (exclusive of noise).

9. Apply 2500°F (Type K) to U.U.T. Adjust potentiometer P11 for a reading

of 2500.0°F.

NOTE: Adjustment steps 8 & 9 are interactive. Repeat these

steps as necessary to get the desired readings.

10. Hit OPR key to store K-meter zero-adjustment in EEPROM. Note that
"FUNC" annuciator turns off.

11. Select Type J thermocouple (KJT key on CL23A, SENSOR SELECT key
on CL24).


12. Turn off U.U.T.

13. Replace K calibration-cable with J calibration-cable.

14. Turn on U.U.T. Note that "FUNC" annunciator is on again. Allow 2-3 min-

utes for thermal restabilization.

15. Apply 32°F (Type J). Adjust P1 for a reading of 32.0°F (exclusive of
noise).

16. Hit OPR key to store J-meter zero-adjustment.

17. Select Type T thermocouple.

18. Turn off U.U.T.


19. Replace J calibration-cable with T calibration-cable.

20. Turn on U.U.T. Allow 2-3 minutes for thermal stabilization.

21. Apply 32°F (Type T). Adjust P1 for a reading of 32.0°F (exclusive of noise).

22. Hit OPR key to store T-meter zero-adjustment.

NOTE: Steps 23 thru 28 apply to Model CL24 only. For Model

CL23A, go to step 29 next.