CALIBRATION PROCEDURES

Test Equipment Required:

1. Thermocouple simulator (OMEGA CL25, or equivalent).

2. Microvolt null-meter (Keithley 155, or equivalent)

3. Calibration cover (OMEGA 820-307-4).

4. Calibration cables, T.C. types K, J, T & E (as required) per Figure 1 below.

5. Low-thermal (eg. copper) shorting-link for null-meter.

NOTE: The (2) alloy-copper junctions in the calibration cables

can be either solder-joints (cleaned to remove flux), or crimp-con-

nections. Insulate each junction with heat-shrink tubing, then

wrap them close together in copper-foil to ensure that they main-

tain identical temperatures (ambient).


Ambient Conditions:


Units should be calibrated at an ambient temperature of 23°C ±1°C, with rel-

ative humidity less than 80%. Test equipment and the unit-under-test

(U.U.T.) should have 1 hour to reach thermal equilibrium with ambient-tem-

perature.

Calibration Adjustments:


1. Remove battery from unit under test (U.U.T.). Refer to Battery

Installation/Replacement instructions in the manual.

2. Remove bottom-cover from U.U.T. Refer to Disassembly Instructions.

3. Remove calibration-jumper J1 from printed-circuit board.


NOTE: When installed, J1 protects the contents of EEPROM

locations where zero-offset adjustments are stored. Once J1 is

removed, this data is erased when the U.U.T. is turned on.


4. Install calibration cover in place of bottom-cover.

5. Re-install battery.

6. Connect U.U.T. to test equipment per Figure 2 (page 43) using the K-type calibra-

tion cable.

7. Turn on U.U.T. and allow 2-3 minutes for thermal re-stabilization of the

TC mini-plug and input-jack.


NOTE: • U.U.T. will be in the Type K, °F meter-mode. The FUNC

annunciator is on, indicating that new constants will be

stored in EEPROM.





39