Table 6: Calibration of Calibrator-Mode Resistance Functions
(RTD-Thermistor)
*exclusive of noise
2. CALIBRATION VERIFICATION
Calibration is verified with the same instrument hookups as used for calibration (ie. Figures 12
and 13 for thermocouple functions, Figure 14 for resistance functions).
While one hookup serves to verify all resistance functions (RTD, thermistor,
ohms), a full check-out of thermocouple functions requires a set of calibration
cables for each thermocouple type.
To simplify the verification of thermocouple calibration, software has been
enhanced to reactivate the copper-mode used during calibration. This feature
is available on products that use the following (and later) software revisions: