14 preset programs, Laser assembly instructions – Odyssey Electronics Marine GPS System User Manual
Page 18
3.1.14 Preset Programs
The Odyssey Navigator comes with procedure specific settings installed.
The Procedures are accessed from the Main Menu by depressing
“Procedures.” Procedures may also be accessed through the “Specialty”
category. By accessing in this way, the procedures are sorted according to
speciality, Fig. 3.1.14.
After selecting a procedure, the settings may be manually adjusted to suit
your needs during a clinical procedure. The settings represent a good starting
point and can be adjusted at any time.
Program Screen Name
Tip Preparation
Technique
Mode
Power
Timer
Abscess
initiated
contact
Continuous
1.0 W
set to 0
Apthous Ulcer
un-initiated
non-contact
Pulse
1.4 W
30 seconds
Biopsy
initiated
contact
Continuous
1.0 W
set to 0
Contouring
initiated
contact
Continuous
0.8 W
set to 0
Denture Sore
un-initiated
non-contact
Pulse
2.0 W
30 seconds
Distal Wedge
initiated
contact
Continuous
1.5 W
set to 0
Draining of Abcesses
initiated
contact
Continuous
0.9 W
set to 0
Emergence Profile
initiated
contact
Continuous
0.8 W
set to 0
Expose Teeth
initiated
contact
Continuous
1.0 W
set to 0
Fibroma
initiated
contact
Continuous
1.0 W
set to 0
Flap
initiated
contact
Continuous
1.0 W
set to 0
Frenectomy
initiated
contact
Continuous
0.8 W
set to 0
Gingivectomy
initiated
contact
Continuous
1.0 W
set to 0
Gingivitis
initiated
contact
Continuous
0.4 W
20 seconds
Gingivitis Therapy
un-initiated
contact
Continuous
0.8 W
20 seconds
Gingivoplasty
initiated
contact
Continuous
1.0 W
set to 0
Hemostasis
un-initiated
non-contact
Continuous
1.0 W
set to 0
Herpetic Lesion
un-initiated
non-contact
Pulse
1.6 W
30 seconds
Hypertrophic tissue
initiated
contact
Continuous
1.5 W
set to 0
Implant exposure
initiated
contact
Continuous
1.0 W
set to 0
Lesion Removal
initiated
contact
Continuous
1.0 W
set to 0
Mucocele
initiated
contact
Continuous
0.9 W
set to 0
Operculectomy
initiated
contact
Continuous
1.5 W
set to 0
Ovate Pontic
initiated
contact
Continuous
1.0 W
set to 0
Periimplantitis
un-initiated
contact
Continuous
1.0 W
set to 0
Perio - Chronic
un-initiated
contact
Continuous
0.5 W
30 seconds
Perio - Severe
un-initiated
contact
Continuous
0.7 W
45 seconds
Socket Treatment
un-initiated
non-contact
Continuous
0.8 W
set to 0
Sulcular Debridement
un-initiated
contact
Continuous
0.8 W
30 seconds
Tissue Tag
initiated
contact
Continuous
1.5 W
set to 0
Tissue Welding
initiated
contact
Continuous
0.6 W
set to 0
Tongue Tie
initiated
contact
Continuous
1.0 W
set to 0
Troughing
initiated
contact
Continuous
0.8 W
set to 0
Vestibuloplasty
initiated
contact
Continuous
1.0 W
set to 0
18
T h e L e a d e r i n D i o d e L a s e r s .
LASER ASSEMBLY INSTRUCTIONS
Section 3
Fig 3.1.14