3.1.11 Installation Test

1.

After login, from the Main Menu, touch the “Quick Set Up” Button

2.

The Odyssey Navigator is pre-set to “Continuous Wave” (CW)

3.

Power setting is pre-set to 1.0W

4.

Hold the fiber approximately 2-4 mm away from a piece of
articulating paper.

5.

Touch “Ready” button – laser is in “Ready” mode.
A green checkmark will be displayed, Fig. 3.1.11,
you will be able to see the aiming beam on the articulating paper
at the end of the fiber tip.

6.

Depress the foot pedal to activate the laser.

7.

The articulating paper will begin to develop smoke in 1-2 seconds.

3.1.12 Check of the Emergency Shutdown System

1.

To check the “Laser Stop” system, simply press and release the red “Laser Stop” button while the laser
is activated. The shut down system will take the laser out of active mode.

2.

To activate the laser, depress the “Laser Stop” button and the laser will resume laser emission.

NOTE: The red Laser Stop button should be depressed for the laser to activate.

Remember: Always test fire the laser outside the mouth before using it on a
patient. The doctor or hygienist, the patient and any staff member present in the
operatory should be wearing the appropriate safety eyewear whenever the laser is
being operated. Strict adherence to protocols for safe laser use is essential.