Table A-3. NI 6115 Analog Input Dynamic Characteristics

Input Range

Bandwidth

(MHz)

1

SFDR Typ

(dB)

2

SFDR Max

(dB)

CMRR

(dB)

3

System Noise

(LSB

rms

)

4

±50 V

5.5

78

70

34

0.35

±20 V

4.4

78

70

40

0.45

±10 V

7.2

81

75

46

0.35

±5 V

4.8

81

75

52

0.35

±2 V

4.8

85

75

60

0.45

±1 V

4.4

85

75

66

0.60

±500 mV

4.4

85

75

70

0.80

±200 mV

4.1

81

70

72

1.3

1

–3 dB frequency for input amplitude at 96% of the input range (–0.3 dB)

2

Measured at 100 kHz with twelfth-order bandpass filter after signal source

3

DC to 60 Hz

4

LSB

rms

, including quantization

Table A-4. NI 6120 Analog Input Dynamic Characteristics

Input Range

Bandwidth

(MHz)

1

SFDR Typ

(dB)

2

SFDR Max

(dB)

3

CMRR

(dB)

4

System Noise

(LSB

rms

)

5

±50 V

1.0

95

90

60

1.2

±20 V

1.0

96

90

68

1.2

±10 V

1.0

95

90

76

1.2

±5 V

1.0

95

90

82

1.5

±2 V

1.0

96

90

90

1.7

±1 V

1.0

94

90

95

2.0

±500 mV

1.0

90

85

100

2.2

±200 mV

1.0

85

80

105

2.8

1

–3 dB frequency for input amplitude at 10% of the input range (–20 dB)

2

Measured at 100 kHz with twelfth-order bandpass filter after signal source

3

100% production tested at 100 kHz

4

DC to 60 Hz

5

LSB

rms

, not including quantization