National Instruments NI 6115/6120 User Manual

Page 82

background image

Appendix A

Specifications

© National Instruments Corporation

A-3

NI 6115/6120 User Manual

Table A-1. NI 6115 Analog Input DC Accuracy Information

Nominal

Range

(V)

Absolute Accuracy

Relative Accuracy

% of Reading

Offset

(mV)

Noise +

Quantization (mV)

Temp

Drift

(%/°C)

Absolute

Accuracy at

Full Scale

(±mV)

Resolution (mV)

Full

Scale

24 Hours

1 Year

Single

Pt.

Averaged

Single

Pt.

Averaged

±50

0.346

0.348

33

42

3.6

0.0229

211.0

48

4.8

±20

0.271

0.273

13

17

1.4

0.0229

69.4

19

1.9

±10

0.026

0.028

6.7

8.3

0.72

0.0004

10.22

10

1.0

±5

0.016

0.018

3.4

4.2

0.36

0.0004

4.61

4.8

0.48

±2

0.036

0.038

1.3

1.8

0.16

0.0004

2.26

2.1

0.21

±1

0.043

0.045

0.68

1.1

0.09

0.0004

1.23

1.2

0.12

±0.5

0.058

0.060

0.35

0.69

0.061

0.0004

0.71

0.80

0.080

±0.2

0.103

0.105

0.15

0.43

0.039

0.0004

0.39

0.51

0.051

Table A-2. NI 6120 Analog Input DC Accuracy Information

1

Nominal

Range

(V)

Absolute Accuracy

Relative Accuracy

% of Reading

Offset

(µV)

2,3

Noise + Quantization

(µV)

Temp

Drift

(%/°C)

Absolute

Accuracy

at Full

Scale

(±mV)

Resolution (µV)

Full

Scale

24

Hours

1 Year

Single

Pt.

Averaged

Single

Pt.

Averaged

±50

0.157

0.159

7,799.2

5,621.9

503.5

0.0106

86.92

6,629.9

663.0

±20

0.139

0.141

3,120.7

2,248.7

201.4

0.0106

31.11

2652.0

265.2

±10

0.033

0.034

1,561.1

1,124.4

100.7

0.0006

4.94

1326.0

132.6

±5

0.036

0.37

781.4

562.2

50.4

0.0006

2.61

663.0

66.3

±2

0.039

0.041

344.0

224.9

20.1

0.0006

1.15

265.2

26.5

±1

0.077

0.079

264.4

150.0

13.7

0.0006

1.05

180.8

18.1

±0.5

0.100

0.102

178.8

144.3

13.7

0.0006

0.69

180.8

18.1

±0.2

0.123

0.125

90.8

112.8

11.0

0.0006

0.35

144.7

14.5

1

Accuracies are valid for measurements following an internal calibration. Averaged numbers assume dithering and

averaging of 100 single-channel readings. Measurement accuracies are listed for operational temperatures within ±1 °C
of internal calibration temperature and ±10 °C of external or factory-calibration temperature.

2

The offset might degrade by 3 LSB with filter enabled.

3

The offset might degrade by 1 LSB when sampling above 500 kS/s.