beautypg.com

Ai start trigger signal, Using a digital source, Routing ai start trigger to an output terminal – National Instruments NI 6238 User Manual

Page 48: Ai start trigger signal -22

background image

Chapter 4

Analog Input

NI 6238/6239 User Manual

4-22

ni.com

AI Start Trigger Signal

Use the AI Start Trigger (ai/StartTrigger) signal to begin a measurement
acquisition. A measurement acquisition consists of one or more samples. If
you do not use triggers, begin a measurement with a software command.
After the acquisition begins, configure the acquisition to stop under the
following conditions:

When a certain number of points are sampled (in finite mode)

After a hardware reference trigger (in finite mode)

With a software command (in continuous mode)

An acquisition that uses a start trigger (but not a reference trigger) is
sometimes referred to as a posttriggered acquisition.

Using a Digital Source

To use ai/StartTrigger with a digital source, specify a source and an edge.
The source can be any of the following signals:

Input PFI <0..5>

RTSI <0..7>

Counter n Internal Output

PXI_STAR

Note Refer to the NI 6238/6239 Specifications for the minimum allowable pulse width
and the propagation delay of PFI <0..5>.

The source also can be one of several other internal signals on your DAQ
device. Refer to Device Routing in MAX in the NI-DAQmx Help or the
LabVIEW 8.x Help for more information.

You also can specify whether the measurement acquisition begins on the
rising edge or falling edge of ai/StartTrigger.

Routing AI Start Trigger to an Output Terminal

You can route ai/StartTrigger out to any output PFI <6..9> or RTSI <0..7>
terminal.

The output is an active high pulse.

The device also uses ai/StartTrigger to initiate pretriggered DAQ
operations. In most pretriggered applications, a software trigger generates
ai/StartTrigger. Refer to the

AI Reference Trigger Signal

section for a

This manual is related to the following products: