beautypg.com

Ai convert clock signal, Using an internal source, Ai convert clock signal -18 – National Instruments NI 6238 User Manual

Page 44: Using an internal source -18

background image

Chapter 4

Analog Input

NI 6238/6239 User Manual

4-18

ni.com

RTSI <0..7>

Input PFI <0..5>

PXI_STAR

ai/SampleClockTimebase is not available as an output on the I/O connector.
ai/SampleClockTimebase is divided down to provide one of the possible
sources for ai/SampleClock. You can configure the polarity selection for
ai/SampleClockTimebase as either rising or falling edge.

AI Convert Clock Signal

Caution Setting the conversion rate higher than the maximum rate specified for your
device will result in errors.

Use the AI Convert Clock (ai/ConvertClock) signal to initiate a single A/D
conversion on a single channel. A sample, controlled by the AI Sample
Clock, consists of one or more conversions.

You can specify either an internal or external signal as the source of
ai/ConvertClock. You also can specify whether the measurement sample
begins on the rising edge or falling edge of ai/ConvertClock.

With NI-DAQmx, the driver will choose the fastest conversion rate possible
based on the speed of the A/D converter and add 10 µs of padding between
each channel to allow for adequate settling time. This scheme enables the
channels to approximate simultaneous sampling and still allow for
adequate settling time. If the AI Sample Clock rate is too fast to allow for
this 10 µs of padding, NI-DAQmx will choose the conversion rate so that
the AI Convert Clock pulses are evenly spaced throughout the sample.

To explicitly specify the conversion rate, use the AI Convert Clock Rate
DAQmx Timing
property node or function.

Using an Internal Source

One of the following internal signals can drive ai/ConvertClock:

AI Convert Clock Timebase (divided down)

Counter n Internal Output

A programmable internal counter divides down the AI Convert Clock
Timebase to generate ai/ConvertClock. The counter is started by
ai/SampleClock and continues to count down to zero, produces an
ai/ConvertClock, reloads itself, and repeats the process until the sample is

This manual is related to the following products: