ERROR

POSSIBLE CAUSE

WHAT TO DO

Note: If more than two minutes have passed since you
applied blood to the test strip, repeat the test with a
new test strip.

2. Clean meter contact points.

2. The meter

contact points
are dirty.

1. Do not remove the test strip

until the meter displays the
result.

1. The test strip

was removed
or partially
removed before
the result was
displayed, or

70

MESSAGES

LSC-99-005 CanSS(RR)OB/Eng 1/5/00 4:03 PM Page 70

Pages: 1 … 79 80 81 82 83 … 90
  • wrong Brand
  • wrong Model
  • non readable
See also other documents in the category Lifescan Health and hygiene:
  • OneTouch UltraMini Blood Glucose Modelling System (2 pages)
  • SureStep blood glucose monitoring system (77 pages)
  • Ultra 2 (74 pages)
  • UltraSmart (134 pages)
  • ONE TOUCH BASIC BLOOD GLUCOSE MONITORING SYSTEM (85 pages)
  • ONE TOUCH Profile Blood Glucose Monitoring System (55 pages)
  • OneTouch Profile blood glucose monitor (74 pages)
  • InDuo blood glucose meter (2 pages)
  • SureStep Blood Glucose Monitor (88 pages)
  • One Touch ll (85 pages)
  • OneTouch Basic Basic Blood Glucose Monitoring System (98 pages)
  • Ultra (73 pages)
  • OneTouch (1 page)
  • OneTouch Diabetes Management Software v2.3.1 (173 pages)
  • One Touch Basic Plus Diabetes Monitoring System (78 pages)
  • Pocket Scan Blood Glucose Monitor System (106 pages)
  • In Duo (90 pages)
  • SureStep Pro (117 pages)
Home Brands Popular manuals Recently added
Manuals Directory

© 2012–2025, azmanual.top
All rights reserved.