Appendix 1—Testing methodology and results for speed and throughput
Kodak’s Quality Assurance Team conducted a series of comprehensive, competitive benchmark
tests comparing the Kodak i780 Scanner with ISIS to the Canon DR-X10C Scanner with VRS.
Throughput tests
The intent of these tests was to determine the throughput of each scanner using typical scanner
configurations. All scanners were configured with similar settings and all tests were conducted
with EMC QuickScan Pro v6.5.1 and run on the same PC host. EMC QuickScan Pro was chosen
because of its popularity as a third party application. The EMC QuickScan Pro demonstration
version was used so it would be easier for interested parties to validate results. (See Reference
A for configuration details.) The operator prepared documents prior to testing and fed them into
each scanner as quickly as possible. Five new sets of documents were used for each scanner
during testing. Each test was run for one hour with the total number of pages recorded.
Chart 1—Throughput testing
Test Number
# Pgs
in 1 hr
Average
ppm
i780 ISIS Test #1
7487
124.8
i780 ISIS Test #2
7557
126.0
i780 ISIS Test #3
7491
124.9
i780 ISIS Test #4
7283
121.4
i780 ISIS Test #5
7510
125.2
i780 ISIS Test #6
7528
125.5
DR-X10C VRS Test #1
5075
84.6
DR-X10C VRS Test #2
2390
39.8
DR-X10C VRS Test #3
5391
89.9
DR-X10C VRS Test #4
3477
58.0
DR-X10C VRS Test #5
1719
28.7
DR-X10C VRS Test #6
4009
66.8
DR-X10C ISIS Test #1
3102
51.7
DR-X10C ISIS Test #2
3142
52.4
DR-X10C ISIS Test #3
5829
97.2
DR-X10C ISIS Test #4
2527
42.1
DR-X10C ISIS Test #5
2293
38.2
DR-X10C ISIS Test #6
4516
75.3
Chart 2—Data and differences on number of pages scanned in one hour
i780
DR-X10C
VRS
DR-X10C
ISIS
i780 Increase
Over
DR-X10C VRS
Performance
i780 Increase
Over
DR-X10C ISIS
Performance
Test #1
7487
5075
3102
2412 (47%)
4385 (141%)
Test #2
7557
2390
3142
5167 (217%)
4415 (140%)
Test #3
7491
5391
5829
2100 (38%)
1662 (28%)
Test #4
7283
3477
2527
3806 (109%)
4756 (188%)
Test #5
7510
1719
2293
5791 (336%)
5217 (227%)
Test #6
7528
4009
4516
3519 (88%)
3012 (67%)
See Reference A for configuration details.
Chart 3—Graphic representation of data and differences on number of pages scanned
in one hour
For more information about testing methodology and laboratory test results, click here
Return to Speed and throughput
0
1000
2000
3000
4000
5000
6000
7000
8000
1
2
3
4
5
6
i780 ISIS
DR X10C VRS
DR X10C ISIS
-
-
Pag
es in One Hour
Test #