Appendix 1—Testing methodology and results for speed and throughput

Kodak’s Quality Assurance Team conducted a series of comprehensive, competitive benchmark
tests comparing the Kodak i780 Scanner with ISIS to the Canon DR-X10C Scanner with VRS.

Throughput tests

The intent of these tests was to determine the throughput of each scanner using typical scanner
configurations. All scanners were configured with similar settings and all tests were conducted
with EMC QuickScan Pro v6.5.1 and run on the same PC host. EMC QuickScan Pro was chosen
because of its popularity as a third party application. The EMC QuickScan Pro demonstration
version was used so it would be easier for interested parties to validate results. (See Reference
A for configuration details.) The operator prepared documents prior to testing and fed them into
each scanner as quickly as possible. Five new sets of documents were used for each scanner
during testing. Each test was run for one hour with the total number of pages recorded.

Chart 1—Throughput testing

Test Number

# Pgs

in 1 hr

Average

ppm

i780 ISIS Test #1

7487

124.8

i780 ISIS Test #2

7557

126.0

i780 ISIS Test #3

7491

124.9

i780 ISIS Test #4

7283

121.4

i780 ISIS Test #5

7510

125.2

i780 ISIS Test #6

7528

125.5

DR-X10C VRS Test #1

5075

84.6

DR-X10C VRS Test #2

2390

39.8

DR-X10C VRS Test #3

5391

89.9

DR-X10C VRS Test #4

3477

58.0

DR-X10C VRS Test #5

1719

28.7

DR-X10C VRS Test #6

4009

66.8

DR-X10C ISIS Test #1

3102

51.7

DR-X10C ISIS Test #2

3142

52.4

DR-X10C ISIS Test #3

5829

97.2

DR-X10C ISIS Test #4

2527

42.1

DR-X10C ISIS Test #5

2293

38.2

DR-X10C ISIS Test #6

4516

75.3

Chart 2—Data and differences on number of pages scanned in one hour

i780

DR-X10C

VRS

DR-X10C

ISIS

i780 Increase

Over

DR-X10C VRS

Performance

i780 Increase

Over

DR-X10C ISIS

Performance

Test #1

7487

5075

3102

2412 (47%)

4385 (141%)

Test #2

7557

2390

3142

5167 (217%)

4415 (140%)

Test #3

7491

5391

5829

2100 (38%)

1662 (28%)

Test #4

7283

3477

2527

3806 (109%)

4756 (188%)

Test #5

7510

1719

2293

5791 (336%)

5217 (227%)

Test #6

7528

4009

4516

3519 (88%)

3012 (67%)

See Reference A for configuration details.

Chart 3—Graphic representation of data and differences on number of pages scanned

in one hour

For more information about testing methodology and laboratory test results, click here

Return to Speed and throughput