14 performance tests – SRS Labs SR850 User Manual
Page 232

7-14
Performance Tests
3) For each of the amplitudes listed below, perform steps 3a through 3c.
AC Calibrator Amplitudes
R Output Expand
1.0000 Vrms
1
100.00 mVrms
10
10.000 mVrms
100
4.000 mVrms
250
a) Set the AC calibrator to the amplitude in the table.
b) Press the keys in the following sequence:
Highlight the R expand.
Enter the R Output Expand from the table.
c) Wait for the R reading to stabilize. Record the value of R.
4) This completes the amplitude linearity test. Enter the results of this test in the test record at the end of
this section.