32 remote programming – SRS Labs SR850 User Manual
Page 208

6-32
Remote Programming
LIA STATUS BYTE
bit
name
usage
0
INPUT/RESRV
Set when an INPUT or RESRV overload is
detected.
1
FILTR
Set when a FILTR overload is detected.
2
OUTPT
Set when an OUTPT overload is detected.
3
UNLK
Set when a reference unlock is detected.
4
RANGE
Set when the detection frequency switches
ranges (harmonic x ref. frequency decreases
below 199.21 Hz or increases above
203.12 Hz). Time constants above 30 s and
Synchronous filtering are turned off in the upper
frequency range.
5
TC
Set when the time constant is changed indirect-
ly, either by changing frequency range, dynamic
reserve or filter slope.
6
TRIG
Set when unit is triggered. Only if samples or
scans are triggered.
7
PLOT
Set when a plot is completed.
The LIA Status bits stay set until cleared by reading or by the
❊
CLS command.
ERROR STATUS BYTE
bit
name
usage
0
Prn/Plt Error
Set when an error occurs during printing or
plotting.
1
Backup Error
Set at power up when the battery backup has
failed.
2
RAM Error
Set when the RAM Memory test finds an error.
3
Disk Error
Set when an error occurs during a disk or file
operation.
4
ROM Error
Set when the ROM Memory test finds an error.
5
GPIB Error
Set when GPIB fast data transfer mode aborted.
6
DSP Error
Set when the DSP test finds an error.
7
Math Error
Set when an internal math error occurs.
The Error Status bits stay set until cleared by reading or by the
❊
CLS command.