Amplitude accuracy and flatness, 11 performance tests, Setup – SRS Labs SR850 User Manual
Page 229: Procedure
![background image](/manuals/147723/229/background.png)
7-11
Performance Tests
4. Amplitude Accuracy and Flatness
This test measures the amplitude accuracy and frequency response.
Setup
We will use the frequency synthesizer to provide an accurate frequency and the AC calibrator to provide a
sine wave with an exact amplitude.
Connect the output of the frequency synthesizer to the phase lock input of the calibrator. Connect the output
of the AC calibrator to the A input of the lock-in. Be sure to use the appropriate terminations where required.
Connect the TTL SYNC output of the synthesizer to the Reference Input of the lock-in.
Set the Synthesizer to:
Set the AC Calibrator to:
Function
Sine
Frequency
1 kHz
Frequency
1 kHz
Amplitude
1.000 Vrms
Amplitude
0.5 Vrms
Voltage
Off
Offset
off or 0V
Phase Lock
On
Sweep
off
Sense
Internal
Modulation
none
Procedure
1) {PRESET} (Turn the lock-in off and on with the [
←
] key pressed)
2) Press the keys in the following sequence:
[REF/PHASE]
Use the knob to select External.
Select Rising edge.
[DISPLAY/SCALE]
Highlight the trace number.
Use the knob to select trace 3 (R) for the top display.
[GAIN/TC]
Select 24 dB/oct.
3) Amplitude accuracy is verified at 1 kHz and various sensitivities. For each sensitivity setting in the
table below, perform steps 3a through 3c.
Sensitivity
AC Calibrator Amplitude
1 V
1.0000 Vrms
200 mV
200.00 mVrms
100 mV
100.000 mVrms
20 mV
20.000 mVrms
10 mV
10.000 mVrms