Spectrum Brands MC.31XX User Manual
Page 75
Trigger modes and appendant registers
Channel Trigger
(c) Spectrum GmbH
75
Channel window trigger for long inner signals
The analog input is continuously sampled with the selected
sample rate. The upper and the lower levels define a win-
dow. Every time the signal enters the window from the out-
side, the pulsewidth counter is startet. If the signal leaves
the window before the pulsewidth counter has stopped, no
trigger will be detected.
If the pulsewidth counter stops and the signal is still inside
the window, the triggerevent will be detected.
Channel window trigger for long outer signals
The analog input is continuously sampled with the selected
sample rate. The upper and the lower levels define a win-
dow. Every time the signal leaves the window from the in-
side, the pulsewidth counter is startet. If the signal enters the
window before the pulsewidth counter has stopped, no trig-
ger will be detected.
If the pulsewidth counter stops and the signal is still outside
the window, the triggerevent will be detected.
Register
Value
Direction
set to
Value
SPC_TRIGGERMODE
40000
r/w
TM_CHANNEL
20040
SPC_TRIGGERMODE0
40200
r/w
TM_CHXWINENTER_LP
10041
SPC_HIGHLEVEL0
42000
r/w
Sets the window’s upper level relatively to the channel’s input range.
board dependant
SPC_LOWLEVEL0
42100
r/w
Sets the window’s lower level relatively to the channel’s input range.
board dependant
SPC_PULSEWIDTH
44000
r/w
Set to the desired pulsewidth in samples.
2 to 255
Register
Value
Direction
set to
Value
SPC_TRIGGERMODE
40000
r/w
TM_CHANNEL
20040
SPC_TRIGGERMODE0
40200
r/w
TM_CHXWINLEAVE_LP
10051
SPC_HIGHLEVEL0
42000
r/w
Sets the window’s upper level relatively to the channel’s input range.
board dependant
SPC_LOWLEVEL0
42100
r/w
Sets the window’s lower level relatively to the channel’s input range.
board dependant
SPC_PULSEWIDTH
44000
r/w
Set to the desired pulsewidth in samples.
2 to 255