MC.31xx Manual
Channel Trigger
Trigger modes and appendant registers
Channel pulsewidth trigger for long positive pulses
The analog input is continuously sampled with the selected
sample rate. If the programmed triggerlevel is crossed by
the channel’s signal from lower to higher values (rising
edge) the pulsewidth counter is started. If the signal crosses
the triggerlevel again in the opposite direction within the
the programmed pulsewidth time, no trigger will be detect-
ed. If the pulsewidth counter reaches the programmed
amount of samples, without the signal crossing the trigger-
level in the opposite direction, the triggerevent will be de-
tected.
The pulsewidth trigger modes for long pulses can be used
to prevent the board from triggering on wrong (short) edges
in noisy signals.
Channel pulsewidth trigger for long negative pulses
The analog input is continuously sampled with the selected
sample rate. If the programmed triggerlevel is crossed by
the channel’s signal from higher to lower values (falling
edge) the pulsewidth counter is started. If the signal crosses
the triggerlevel again in the opposite direction within the
the programmed pulsewidth time, no trigger will be detect-
ed. If the pulsewidth counter reaches the programmed
amount of samples, without the signal crossing the trigger-
level in the opposite direction, the triggerevent will be de-
tected.
The pulsewidth trigger modes for long pulses can be used
to prevent the board from triggering on wrong (short) edges
in noisy signals.
Register
Value
Direction
set to
Value
SPC_TRIGGERMODE
40000
r/w
TM_CHANNEL
20040
SPC_TRIGGERMODE0
40200
r/w
TM_CHXPOS_LP
10001
SPC_HIGHLEVEL0
42000
r/w
Set it to the desired triggerlevel relatively to the channel’s input range.
board dependant
SPC_PULSEWIDTH
44000
r/w
Set to the desired pulsewidth in samples.
2 to 255
Register
Value
Direction
set to
Value
SPC_TRIGGERMODE
40000
r/w
TM_CHANNEL
20040
SPC_TRIGGERMODE0
40200
r/w
TM_CHXNEG_LP
10011
SPC_HIGHLEVEL0
42000
r/w
Set it to the desired triggerlevel relatively to the channel’s input range.
board dependant
SPC_PULSEWIDTH
44000
r/w
Set to the desired pulsewidth in samples.
2 to 255