Spectrum Brands MC.31XX User Manual
Page 72
72
MC.31xx Manual
Channel Trigger
Trigger modes and appendant registers
Channel steepness trigger for flat positive pulses
The analog input is continuously sampled with the selected
sample rate. If the programmed lower level is crossed by
the channel’s signal from lower to higher values (rising
edge) the pulsewidth counter is started. If the signal does
cross the upper level within the the programmed pulsewidth
time, no trigger will be detected.
If the pulsewidth counter reaches the programmed amount
of samples a triggerevent will be detected.
Channel steepness trigger for flat negative pulses
The analog input is continuously sampled with the selected
sample rate. If the programmed upper level is crossed by
the channel’s signal from higher to lower values (falling
edge) the pulsewidth counter is started. If the signal does
cross the lower level within the the programmed pulsewidth
time, no trigger will be detected.
If the pulsewidth counter reaches the programmed amount
of samples a triggerevent will be detected.
Register
Value
Direction
set to
Value
SPC_TRIGGERMODE
40000
r/w
TM_CHANNEL
20040
SPC_TRIGGERMODE0
40200
r/w
TM_CHXPOS_GS
10003
SPC_HIGHLEVEL0
42000
r/w
Set it to the desired upper level relatively to the channel’s input range.
board dependant
SPC_LOWLEVEL0
42100
r/w
Set it to the desired lower level relatively to the channel’s input range.
board dependant
SPC_PULSEWIDTH
44000
r/w
Set to the desired pulsewidth in samples.
2 to 255
Register
Value
Direction
set to
Value
SPC_TRIGGERMODE
40000
r/w
TM_CHANNEL
20040
SPC_TRIGGERMODE0
40200
r/w
TM_CHXNEG_GS
10013
SPC_HIGHLEVEL0
42000
r/w
Set it to the desired upper level relatively to the channel’s input range.
board dependant
SPC_LOWLEVEL0
42100
r/w
Set it to the desired lower level relatively to the channel’s input range.
board dependant
SPC_PULSEWIDTH
44000
r/w
Set to the desired pulsewidth in samples.
2 to 255