Fluke ESA612 User Manual
Page 77

Electrical Safety Analyzer
Detailed Specifications
63
Ranges................................................................0.0 to 199.9
μ
A
200 to 1999
μ
A
2.00 to 10.00 mA
Accuracy
DC to 1 kHz.....................................................
±
(1 % of reading + (1
μ
A or 1 LSD, whichever is greater))
1 to 100 kHz ....................................................
±
(2 % of reading + (1
μ
A or 1 LSD, whichever is greater))
100 kHz to 1 MHz............................................
±
(5 % of reading + (1
μ
A or 1 LSD, whichever is greater))
Mains on applied part test voltage ......................100 %
±
5 % of Mains for AAMI, current limited to 1 mA
±
25 % per AAMI
100
%
±
5 % of Mains for 62353 current limited to 3.5 mA
±
25 % per 62353
100
%
±
5 % of Mains for 60601 current limited to 7.5 mA
±
25 % per 60601
Differential leakage
Ranges................................................................10 to 199
μ
A
200 to 2000
μ
A
2.00 to 20.00 mA
Accuracy..............................................................
±
(10 % of reading + (2 counts or 20
μ
A, whichever is greater))
Insulation resistance
Ranges................................................................0.5 to 20.0 M
Ω
20.0 to 100.0 M
Ω
Accuracy
20 M
Ω
Range..................................................
±
(2 % of reading + 0.2 M
Ω
)
100 M
Ω
Range................................................
±
(7.5 % of reading + 0.2 M
Ω
)
Source test voltage .............................................500 V dc (+20 %, -0 %) 1.5
±
0.25 mA short-circuit current or 250 V dc selectable
Maximum load capacitance.................................1
μ
F