Introduction, Electrical safety analyzer – Fluke ESA612 User Manual
Page 15
1
Electrical Safety Analyzer
Introduction
The Fluke Biomedical ESA612 Electrical Safety Analyzer
(hereafter the Analyzer) is a full-featured, compact,
portable analyzer, designed to verify the electrical safety
of medical devices. The Analyzer tests to domestic
(ANSI/AAMI ES1, NFPA 99) and international (IEC62353,
AN/NZS 3551, and parts of IEC 60601-1) electrical-safety
standards. The integrated ANSI/AAMI ES1 and
IEC60601-1 patient loads are easily selectable.
The Analyzer performs the following tests:
•
Line (Mains) voltage
•
Ground Wire (or Protective Earth) Resistance
•
Equipment
current
•
Insulation
resistance
•
Ground (Earth) leakage
•
Chassis (Enclosure) leakage
•
Lead to Ground (Patient) and Lead to Lead (Patient
Auxiliary) leakage
•
Lead isolation (Mains on applied parts leakage)
•
Differential
leakage
•
Direct equipment leakage
•
Direct applied part leakage
•
Alternative equipment leakage
•
Alternative applied part patient leakage
•
Point to point leakage, voltage, and resistance
•
ECG simulation and performance waveforms