Performing a current consumption test, Performing leakage current tests – Fluke ESA612 User Manual
Page 39
Electrical Safety Analyzer
Performing Electrical Safety Tests
25
Performing a Current Consumption Test
To measure the current consumed by the DUT, press
A
. The Analyzer displays the current flowing through
the mains connections of the test receptacle.
Performing Leakage Current Tests
The Analyzer measures leakage current for a number of
different DUT configurations. In addition to the leakage
found on the enclosure and the earth connection, the
Analyzer can measure leakage on each connected applied
part and combinations of connected applied parts.
Which leakage tests are available depends on which
standard is selected. See the Selecting the Test Standard
section earlier in this manual to change the standard the
Analyzer is using.
Table 5 lists six leakage current tests that have different
names based on which standard is selected.
Press B to access the leakage current main menu
shown in Figure 17.
Table 5. Test Names Based on Selected Standard
IEC60601 AAMI/NFPA
99
Protective Earth Resistance
Ground Wire Resistance
Earth Leakage Current
Ground Wire Leakage Current
Touch or Enclosure Leakage Current
Chassis Leakage Current
Patient Leakage Current
Lead to Ground Leakage Current
Patient Auxiliary Leakage Current
Lead to Lead Leakage Current
Mains on Applied Part (MAP) Leakage Current
Isolation Leakage Current