Fluke ESA612 User Manual
Page 62
ESA612
Users Manual
48
When performing an applied parts test using the
AAMI/NFPA-99 standard, the normal connections of RA,
LL, LA, and RL are made to their associated Analyzer
input jacks. Four adapters from the Universal Snap to
Banana Adapter set will be required for the first four
connections. The remaining chest leads are connected to
the Adapter and the common lead from the Adapter is
plugged into the V1 jack (5
th
jack) of the Analyzer. See
Figure. 30. This configuration allows for isolation of the
RA, LL, LA, and RL leads from each other and the
remaining chest leads, which are shorted together, while
performing leakage tests.