beautypg.com

Using the 1210 adapter – Fluke ESA612 User Manual

Page 60

background image

ESA612
Users Manual

46

Using the 1210 Adapter

The 1210 Adapter, an optional accessory, is designed to
increase the number of lead or applied parts connections
to the Analyzer from five to 14. The adapter itself ties up to
ten leads together into a single lead that is plugged into
one of the input jacks of the Analyzer. The four remaining
Analyzer input jacks can also be used in conjunction with
the Adapter.
Figure 29 shows one application of the Adapter. The
Defibrillator/Monitor in the example has ten ECG leads,
two pacer leads, and two defibrillator paddles which need
to be tested together, and in groups if single function, for
current leakage per IEC62353. The example shows the
ECG leads to be snap type connectors and two BJ2ECG
adapters are shown plugged into the Adapter. If the ECG
leads did not have snap connectors, then the Universal
Snap to Banana Adapter can be used to make the
connections to the Adapter. The common lead from the
Adapter is plugged into the RA jack (1

st

jack) of the

Analyzer. Using four sheathed test leads with alligator
clips, connect the two defibrillator paddles into the LL and
LA Analyzer jacks and the two pacer leads into the RL and
V1 jacks. Using the selection that ties all five Analyzer
jacks together will test all fourteen leads for leakage
current. Using the selection of applied part group of 1, 2,
and 2 allows for testing of groups of applied parts of single
function.