Detailed Specifications

Test Standard Selections........................................ ANSI/AAMI ES-1, IEC62353, IEC60601-1, and AN/NZS 3551
Voltage

Ranges (Mains voltage)...................................... 90.0 to 132.0 V ac rms

180.0 to 264.0 V ac rms

Range (Point-to-point voltage) ............................ 0.0 to 300.0 V ac rms
Accuracy .............................................................

±

(2 % of reading + 0.2 V)

Earth Resistance

Modes ................................................................. Two terminal
Test Current ........................................................ >200 mA ac
Range ................................................................. 0.000 to 2.000

Ω

Accuracy .............................................................

±

(2 % of reading + 0.015

Ω

)

Equipment Current

Range ................................................................. 0.0 to 20.0 A ac rms
Accuracy .............................................................

±

(5 % of reading + (2 counts or 0.2A, whichever is greater))

Duty cycle ........................................................... 15 A to 20 A, 5 min. on/5 min. off

10 A to 15 A, 7 min. on/3 min. off

0 A to 10 A, continuous

Leakage Current

Modes*................................................................ AC+DC (True-rms)
AC

only

DC

only

* Modes: AC+DC, AC only, and DC only available for all leakages with exception of MAP
that are available in True-rms (shown as AC+DC)

Patient Load Selection........................................ AAMI ES1-1993 Fig. 1

IEC 60601: Fig. 15

Crest factor .........................................................

3