Operating the Meter Using the Computer Interface

Computer Interface Command Set

5

5-25

Range and Measurement Rate Commands and Queries

The commands in Table 5-13 relate to ranging and measurement rates (i.e., readings/
second). In the autorange mode, the meter automatically selects a range for each reading;
in the manual range mode, the user selects a fixed range.

Table 5-13. Range and Measurement Rate Commands and Queries

Command

Description

Causes the meter to enter the autoranging mode on the primary display. If the
autorange mode cannot be selected (e.g., if REL, dB, MN MX, or diode/continuity test
is selected), an Execution Error is generated.

Causes meter to return "1" if it is in autorange, or "0" if it is not.

Causes the meter to exit autoranging on the primary display and enter manual ranging.
The present range becomes the selected range.

Sets the primary display to the range designated by . is a number
between 1 and 7 that corresponds to a range shown in Tables 5-12A and 5-12B.

Table 5-13A. Ranges at Fast & Medium Measurement Rate

Range

Value

Voltage

Range

Ohms

Range

Current

Range

Frequency

Range

1
2
3
4
5
6
7

300 mV

3 V

30 V

300 V

1000 V dc*

ERROR
ERROR

300

3 k

30 k

300 k

3 M

30 M

300 M

30 mA

100 mA

10 A

ERROR
ERROR
ERROR
ERROR

1000 Hz

10 kHz

100 kHz

1000 kHz

1 MHz

ERROR
ERROR

* 1000 V dc, 750 V ac

Table 5-13B. Ranges at Slow Measurement Rate

Range

Value

Voltage

Range

Ohms

Range

Current

Range

Frequency

Range

1
2
3
4
5
6
7

100 mV

1000 mV

10 V

100 V

1000 V dc*

ERROR
ERROR

100

1000

10 k

100 k

1000 k

10 M

100 M

10 mA

100 mA

10 A

ERROR
ERROR
ERROR
ERROR

1000 Hz

10 kHz

100 kHz

1000 kHz

1 MHz

ERROR
ERROR

* 1000 V dc, 750 V ac

AUTO

AUTO?

FIXED

RANGE