Ohms, Diode test/continuity – Fluke 45 User Manual
Page 111

Appendices
Specifications
A
A-7
Ohms
Resolution
Range
Slow
Medium
Fast
Accuracy
Typical Full
Scale
Voltage
Max Current
Through the
Unknown
300
Ω
3 k
Ω
30 k
Ω
300 k
Ω
3 M
Ω
30 M
Ω
300 M
Ω
*
100
Ω
1000
Ω
10 k
Ω
100 k
Ω
1000 k
Ω
10 M
Ω
100 M
Ω
*
—
—
—
—
—
—
—
1 m
Ω
10 m
Ω
100 m
Ω
1
Ω
10
Ω
100
Ω
100 k
Ω
10 m
Ω
100 M
Ω
1
Ω
10
Ω
100
Ω
1 k
Ω
100 k
Ω
—
—
—
—
—
—
—
100 M
Ω
1
Ω
10
Ω
100
Ω
1 k
Ω
10 k
Ω
1 M
Ω
—
—
—
—
—
—
—
0.05 % + 2 + 0.02
Ω
0.05 % + 2
0.05 % + 2
0.05 % + 2
0.06 % + 2
0.25 % + 3
2 %
0.05 % + 8 + 0.02
Ω
0.05 % + 8 + 0.02
Ω
0.05 %+8
0.05 % + 8
0.06 % + 8
0.25 % + 6
2 % + 2
0.25
0.24
0.29
0.29
0.3
2.25
2.9
0.09
0.10
0.11
0.11
0.12
1.5
2.75
1 mA
120µA
14 µA
1.5 µA
150 µA
320 µA
320 µA
1 mA
120 µA
14 µA
1.5 µA
150 µA
150 µA
320 µA
*Because of the method used to measure resistance, the 100 M
Ω
(slow) and 300 M
Ω
(medium and fast)
ranges cannot measure below 3.2 M
Ω
and 20 M
Ω
, respectively. "UL" (underload) is shown on the display for
resistances below these nominal points, and the computer interface outputs "+1 E-9".
Open Circuit Voltage
3.2 V maximum on the 100
Ω
, 300
Ω
, 30 M
Ω
, 100 M
Ω
, and 300 M
Ω
ranges, 1.5 V
maximum on all other ranges.
Input Protection
500 V dc or rms ac on all ranges
Diode Test/Continuity
Maximum Reading
Resolution
Slow
Medium
Fast
999.99 mV
2.5 V
2.5 V
10 µV
100 µV
1 mV
Test Current
Approximately 0.7 mA when measuring a forward biased junction.
Audible Tone
Continuous tone for continuity. Brief tone for normal forward biased diode or
semiconductor junction.