Fluke 45 User Manual
Page 73

45
Users Manual
5-14
the bus responds to the poll by sending the contents of its Status Byte Register. If an
instrument on the bus has made a service request, the request service bit (RQS, bit 6) of its
Status Byte Register will be set to 1, identifying it as an instrument that requested service.
The contents of the Status Byte Register (STB) is determined by the Service Request
Enable Register (SRE), Event Status Register (ESR), Event Status Enable Register
(ESE), and the output buffer. These status registers are discussed below and summarized
in Table 5-5. Figure 5-3 shows the relationship of these registers.
Table 5-5. Status Register Summary
Register
Read Command
Write Command
Enable Register
Status Byte Register
Service Request Enable Register
Event Status Register
Event Status Enable Register
*STB?
*SRE?
*ESR?
*ESE?
None
*SRE
None
*ESE
SRE
None
ESE
None