Table 5-12. Function Modifier Commands and Queries (cont.)
Command
Description
MAXSET
MIN
MINSET
MMCLR
MOD?
REL
RELCLR
RELSET
RELSET?
Meter enters MN MX modifier with
number with exponent. Autoranging is disabled. See "MN MX Modifier" in Chapter 3.
If
Meter enters MN MX modifier with present reading as minimum value. If already in
MN MX modifier, meter displays minimum value. In MN MX modifier, autoranging is
disabled. See "MN MX Modifier" in Chapter 3.
Meter enters MN MX modifier with
number with exponent. Autoranging is disabled. See "MN MX Modifier" in Chapter 3.
If
Meter exits the MN MX modifier. The stored minimum and maximum values are lost,
and the meter returns to the ranging mode and range selected prior to selecting MN
MX modifier.
Note
When the MN MX modifier is selected, you can toggle
between displaying the minimum and maximum reading
without losing the minimum and maximum values stored.
Meter returns a numeric value indicating modifiers in use. 1 = MN, 2 = MX, 4=HOLD,
8=dB, 16=dB Power, 32=REL, 64=COMP.
If multiple modifiers are selected, the value returned is equal to the sum of the values of
the selected modifiers. For example, if dB and REL are selected, "40" is returned.
Meter enters the relative (REL) modifier, using the value shown on the primary display
as the relative base. Autoranging is disabled. See "REL Modifier" in Chapter 3 for more
on the relative modifier.
Meter exits the relative modifier and returns to the ranging mode and range selected
prior to selecting the relative modifier.
Meter enters the relative (REL) modifier, using
signed real number with exponent. Autoranging is disabled.
If
See "REL Modifier" in Chapter 3 for more on the relative modifier.
Meter returns
Execution Error is generated.