Fluke 45 User Manual
Page 34

Operating the Meter From the Front Panel
Ranging
3
3-7
Table 3-3. Current Ranges and Full Scale Values
Fast Reading Rate
Medium Reading Rate
Slow Reading Rate
Range
Full Scale
Range
Full Scale
Range
Full Scale
30 mA
30.00 mA
30 mA
30.000 mA
10 mA
9.9999 mA
100 mA
100.0 mA
100 mA
100.00 mA
100 mA
99.999 mA
10 A
10.00 A*
10 A
10.000 A*
10 A
9.9999 A
* 20 A for maximum of 30 seconds
Table 3-4. Ohms Ranges and Full Scale Values
Fast Reading Rate
Medium Reading Rate
Slow Reading Rate
Range
Full Scale
Range
Full Scale
Range
Full Scale*
300
Ω
300.0
Ω
300
Ω
300.00
Ω
100
Ω
98.000
Ω
3 k
Ω
3.000 k
Ω
3 k
Ω
3.0000 k
Ω
1000
Ω
980.00
Ω
30 k
Ω
30.00 k
Ω
30 k
Ω
30.000 k
Ω
10 k
Ω
9.8000 k
Ω
300 k
Ω
300.0 k
Ω
300 k
Ω
300.00k
Ω
100 k
Ω
98.000 k
Ω
3 M
Ω
3.000 M
Ω
3 M
Ω
3.0000 M
Ω
1000 k
Ω
980.00 k
Ω
30 M
Ω
30.00 M
Ω
30 M
Ω
30.000 M
Ω
10 M
Ω
9.8000 M
Ω
300 M
Ω
300 M
Ω
300 M
Ω
300.0 M
Ω
100 M
Ω
98.0 M
Ω
**
*Typical
** Because of the method used to measure resistance, the 100 M (slow) and 300 M (medium and fast)
ranges cannot measure below 3.125 M and 20 M respectively. "UL" (Underload) is shown on the display
for resistances below these nominal points, and the computer interface outputs "+1 E-9".
Table 3-5. Frequency Ranges and Full Scale Values (Slow and Medium*)
Range
Full Scale
1000 Hz
999.99 Hz
10 kHz
9.9999 kHz
100 kHz
99 999 kHz
1000 kHz
999.99 kHz
1 MHz
9.9999 MHz
* Fast (F) reading rate has one digit of resolution less.