1 fill to setpoint – Industrial Data Systems IDS 440 User Manual
Page 17
IDS440 Users Manual version 4.2
Built In Application Programs
13
#1 Fill to Setpoint
Press F1 key to enter a setpoint. Place an empty container on the scale. Press F2 key to
begin filling. TTL output 1 turns on until the weight on the scale is equal or greater than
the setpoint.
Installation
Connect TTL output 1 to relay or lamp actuator.
TB4 - Pin 1 TTL output 1
TB4 - Pin 16 Ground (GND)
Setup Parameters:
(See chapter 5: Set Operation Parameters for setup details)
Parameter 43 Minimum scale weight to begin filling. Verifies that a container is on the
scale. Set to 0 if there is no minimum.
Parameter 44 Minimum fill amount. Set to some small value to insure that a fill amount
has been entered.
Parameter 45 Maximum fill amount. Maximum fill weight that will ever be used.
Parameter 46 Scale empty weight. Maximum weight to begin fill. Verifies that an empty
container is on the scale.
Operators Functions
Enter setpoint
Press
F1 key to enter the setpoint. Enter the setpoint weight, then press the ENTER key.
Start fill operation
Place the container on the scale (if required). Press
F2 key to begin fill operation.
Stop fill operations
Press the
Clear key to stop fill operation.
Print Results
Press the Print key to print the Gross, Tare, and Net results of the fill operation.
Error Codes
Error 30 = scale not empty
Error 31 = minimum weight error: container is not on scale
Error 32 = tare weight error: tare function failed, scale not stable (motion)
Error 33 = setpoint less than minimum setpoint amount
Error 34 = setpoint more than maximum setpoint amount