Install the specimen – MTS Landmark TWX User Manual
Page 77

a) Similarly, open the Tuning: Ch1 Force tab by selecting Station Manager > Display > Station
Setup > Channels > Ch1 (or other depending on your configuration) > Force.
b) Use the following parameters for this hard specimen test:
Important: The following numbers were appropriate for the example test on a typical system.
You will need to adapt these entries to your specimen, test, and system.
•
P Gain — 1
•
I Gain — 1
•
D Gain — 0
•
F Gain — 0
•
F2 Gain — 0
•
Proportional FF Gain — 0
•
FL Filter — Set to maximum positive value
4. Retain the Tuning user access level.
Leave the Operator Type selection set to Tuning because you will do more tuning once the specimen
is installed.
Install the Specimen
Installing the specimen consists of the following tasks:
a)
(p. 78)
b)
Position the Actuator Using Displacement Control
(p. 79)
MTS Landmark Test System Operation | 77