beautypg.com

MTS Landmark TWX User Manual

Page 113

background image

a) Navigate to Station Manager > Display > Station Setup > Channels > Ch1 (or other depending

on your configuration) > Force > Limits.

b) Enter -0.05 kN in the Lower Limit text box and ensure that Interlock is selected for the Action.

8. Release the upper grip.

Using the controls on the load frame, release the upper grip.

9. If necessary, repeat the upper grip clamping process.

If Necessary, Recover from a Tripped Displacement Limit

There are a number of possible causes for a tripped displacement limit. For example:

The actuator may have retracted below the lower displacement limit during specimen installation.

The actuator may have retracted when the specimen broke at the conclusion of the test.

Your test limits may have been insufficient for the test.

Rather than cover all possibilities, this section addresses the first case only. Adapt this procedure as
necessary to recover from other situations.

MTS Landmark Test System Operation | 113

Running the Example Tension Test